Cryopump Repair Service
 |
- Vacuum Side
- Carbon array and indium are replaced
- 80K array is given a new nickel plate
- Radiation shield is refinished
- Parts are cleaned and baked out
- He leak checked to 1x10-8 cc/s
|
- Helium Circuit
- Bearings, O-rings, seals and springs are replaced
- 1st and 2nd stage displacers cleaned and baked out
- Displacer motor is regaussed
 | - Testing
- Pump is charged with ultra pure helium and tested over a 24 hour period
- Regeneration cycle is run to test cool down time
- On-Board electronics are tested
- Pumps are tagged to reflect work done and test data
|
Cryo Compressor Rebuilds
Cryo compressors are also rebuilt to exacting standards. We perform the following:
- Replacement of valves and adsorber
- Replacement of marginal mechanical and electronic parts (extra charge, requires customer approval)
- Change oil
- Recharge with ultra pure helium
- Clean helium lines
- Helium leak sniff compressor, lines and fittings
Service Index e-mail PTB home